Model
|
ENDA-5000
|
|
Component
|
NOx
|
SO2
|
CO
|
CO2
|
O2*1
|
|
Measurement methods
|
NDIR
|
NDIR
|
NDIR
|
NDIR
|
Magneto-pneumatic detection
|
|
Range*2
|
Standard
|
200~5000 ppm
|
200~5000 ppm
|
200~5000 ppm
|
5~25 vol%
|
10~25 vol%
|
|
|
Optional
|
100 ppm~
|
50 ppm~
|
100 ppm~
|
-
|
-
|
|
Range Ratio
|
Within a factor of 10
|
Within a factor of 10
|
Within a factor of 10
|
Within a factor of 5
|
Within a factor of 2.5
|
|
Repeatability
|
Within 0.5% of full scale (with optional range, or during O2 measurement, ±1.0% of full scale)
|
|
Linearity (indicator error)
|
±1.0% of full scale
|
|
Zero drift
|
±1.0% of full scale (assuming surrounding temperature is maintained within 5 °C) (with optional range, or O2 measurement, ±2.0% of full scale)
|
|
Span drift
|
±2.0% of full scale/week (assuming surrounding temperature is maintained within 5°)
|
|
Response time*3
|
Within 60 seconds (Td + T90 from equipment intake area) (sample flow 0.6 L/min.) (within 240 seconds for SO2 only)
|
|
Overall interference frrom co-existing gases
|
±2.0% of full scale/week (within standard range, with standard gas formation)
|
|
Display
|
Touch panel LCD (backlight) (four usable lines)
|
|
Environment Condition
|
Temperature
|
-5 to 40 °C (away from direct sunlight and radiation heat )*4
|
|
Humidity
|
85% or less (no condensation)
|
|
Vibration
|
100 Hz, 0.3 m/S2 or less
|
|
Dust
|
Standard environment or better
|
|
Measuring Gas Condition
|
Temperature
|
250 °C or lower
|
|
Dust
|
0.1 g/Nm3 or less
|
|
Standard gas composition*5
|
NO: 500 ppm or less; NO2: 15 ppm or less; SO2: 1000 ppm or less; SO3: 50 ppm or less; CO: 200 ppm or less; CO2: 15% volume or less; H2O: 40% volume or less
|
|
Sampling method
|
Dry sampling using an electric cooler
|
|
Sample gas flow
|
2.5 L/min~3.0 L/min
|
|
Sample inlet tube
|
Teflon tubing (Ø8/Ø6 mm)
|
|
Sample gas pressure
|
±4.9 kPa (three points selected) (with no sample gas back pressure)
|
|
(1) -1.96 to 4.9 kPa (2) ±3.43 kPa (3) -4.9 to 1.96 kPa
|
|
|
Pressure control
|
Pressure control uses a regulator and cylinder; Reduced pressure sampling; Control pressure: -4.9 kPa
|
|
Output
|
DC 4 to 20 mA (absolute output) (DC 0 to 16 mA/DC 0 to 1 V/DC 1 to 5V optional) Max. 12 output systems
|
|
External output
|
Analysis alerts, analysis warnings, range display, corrections, conservation, purging (option) Contact capacity: DC 30 V 1 A, AC 250 V 1 A resistance load
|
|
Correction method
|
Dry correction, automatic correction (correction cycle: 7 days standard, can be adjusted to between 1 and 99 days), manual correction
|
|
Calibration gas
|
Zero gas
|
With measurement method authorization: N2, When there is no measurement method authorization: N2 or ambient air
|
|
O2carrier gas
|
Ambient air
|
|
Span gas
|
Gas cylinder for each component measured (when there is no measurement method authorization: O2 or ambient air can be used)
|
|
Probe collection point filter
|
Flange: JIS 10K, 40 AFF; Sample probe tube length: 1000 mm; Material: SUS-316 stainless steel; Filter element: SUS-304 stainless steel and 2um-pleated quartz wool; Electric heater: 100 VA, with water droplet proof case
|
|
Power supply
|
AC 100 V ±15 V(85 V~115 V)
|
|
Power frequency
|
50/60 Hz (switchable)
|
|
Power consumption
|
About 800 VA (heated piping 30m: +1100 VA; heater in tray: +300 VA)
|
|
Exterior dimensions/Mass
|
600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L cylinders, maximum of 3 cylinders); About 180 kg (not including cylinders) 600 (W) x 1770 (H) x 300 (D) mm (high pressure gas cylinders, 3.4 L cylinders, maximum of 3 cylinders); About 200 kg (not including cylinders)
|
|
Materials in contract with sample gas
|
SUS-316 stainless steel, SUS-304 stainless steel, Teflon, polypropylene, polyethylene, flourorubber, PVC, PVDF, and glass
|
|
Enclosure
|
Independent outdoor installation Plate thickness: Main unit, doors, top plate, steel plate: 2.3 mm; Channel base: 3.2 mm; Doors: front opening; Interface: right front
|
|
Color/Finish
|
Semi-gloss Munsell 5Y7/1 on all inner and outer surfaces
|